Digikey Voltage Box: (2.5hr soak, 64F ambient, 100PLC, HighZ input)
Reading Scale Other 1 yr Drift (Vref + DMM)
100.00097mV 100mV Null + Avg 20 samples 1.57 uV
0.9999591V 1V Null + Avg 20 samples -51 uV
2.499842V 10V Avg 20 samples 82 uV
4.999121V 10V Avg 20 samples -279 uV
10.000018V 10V Avg 20 samples -12 uV
AliExpress Box: (6hr soak, 62F ambient, 100PLC, HighZ input)
Reading Scale Other 1 yr Drift (Vref + DMM)
99.99982mV 100mV Null + Avg 20 Samples -1.99 uV
1.0000619V 1V Null + Avg 20 Samples 50.9 uV
2.500164V 10V Avg 20 Samples 24 uV
10.000056V 10V Avg 20 Samples 26 uV
Current Reference:
64F
Value Scale 1 yr Drift (Iref + DMM)
100.00127uA 100uA Avg 20 samples 8.14 nA
0.9999867mA 1mA Avg 20 samples 9.7 nA
9.999337mA 10mA Avg 20 samples -2.137 uA
Conclusions:
After 2 years the meter and references are well within the expected tolerances for both the voltage and current ranges.