It's been almost 20 months since the last measurement of the V2 reference. Here's the data:
Temperature = 62F (~16.7C), HighZ input, 100plc
Reading Scale Other Error Drift (Since last test)
99.98991mV 100mV Null, Avg 20 samples ~-10.1uV -2.15uV
1.0000764V 1V Avg 20 samples +76uV -6.4uV
10.00000V 10V Avg 20 samples <5uV -70uV
Can't complain about these results. They aren't indicative of any error or drift in either instrument.
The DVM soaked for a bit more than 1 hr before I took the measurements.