Same conditions as last year:
Temperature = 70F (21C), HighZ input, 100plc
Reading Scale Other Error Drift (Since last test)
99.99250 mV 100mV Null, Avg 20 samples ~-7.5uV +9.34uV
1.000077 V 1V Null, Avg 20 samples +76uV +6uV
10.000036V 10V Avg 20 samples +36uV +36uV
Everything is still looking pretty good.